Inlet or detector pressure error

Short checks only. This is not a repair procedure. Stop for a smell of gas, a hydrogen or leak shutdown, a damaged fitting or column, or a fault that does not change.

Quick checks

Cause

Inlet: gas off, septum or ferrule leak, column broken or not installed, plugged split vent, or column length and inner diameter in the method that do not match the column. Detector: makeup gas, or FID hydrogen or air, depending on the message.

Check

Read the message. Do not swap gases to test the fault.

When to call

If you smell gas or a leak shutdown trips, ventilate and call. Never use a flame. Hydrogen in an oven is a fire hazard.

Full notes

An inlet pressure fault usually means the instrument cannot hold the setpoint. Routine causes are the gas supply turned down or empty, a septum leak, a column ferrule leak, a column that is broken or not installed, a plugged split vent trap, or column dimensions in the method that do not match the column in the oven. Electronic pressure control calculates flow from those dimensions. Wrong length or inner diameter produces the wrong pressure.

A detector pressure fault points at makeup gas, FID hydrogen or air, or the detector gas fitting, depending on the detector. Read the message. Do not swap gases to “see what happens.”

If the system shuts down on a leak, or if you smell gas, ventilate, stop, and call Mahut. Never use a flame to find a leak. Hydrogen collecting in an oven or a closed room is a fire hazard.